Comparison of GeneXpert MRSA/SA ETA assay with semi-quantitative and quantitative cultures and nuc gene-based qPCR for detection of Staphylococcus aureus in endotracheal aspirate samples
Antimicrobial Resistance & Infection Control20198:4
© The Author(s). 2019
- Received: 2 October 2018
- Accepted: 21 December 2018
- Published: 5 January 2019
Abstract
Introduction
Staphylococcus aureus (S. aureus) is a common cause of ventilator-associated pneumonia. Rapid and accurate detection of lower respiratory tract colonization and/or infection with S. aureus may inform targeted preventive and therapeutic strategies. To investigate this, we compared semi-quantitative (SQ)-culture results from 79 endotracheal aspirates (ETA) collected from mechanically-ventilated patients, to two culture and two non-culture-based methods for detection of S. aureus.
Methods
ETA analyzed by routine SQ-culture on blood and colistin-nalidixic-acid agar was compared to: (i) quantitative (Q-) culture on chromogenic COLOREX™ Staph aureus; (ii) enrichment in brain-heart-infusion broth followed by plating on blood agar and COLOREX™; (iii) nuc-based TaqMan qPCR, and (iv) GeneXpert MRSA/SA ETA assay.
Results
Of the 79 ETA samples analyzed by SQ-culture, 39 samples were positive, and 40 negative for S. aureus. Two samples negative for S. aureus by SQ-culture were, however, S. aureus-positive by the other four methods and were considered positive. Appending these two samples as positive in the SQ-culture results, sensitivities−specificities for Q-culture, enrichment-culture, TaqMan qPCR and GeneXpert were 100–95, 100–92, 100–53% and 100% − 100, respectively. The lower specificities of Q-culture, enrichment-culture, and TaqMan qPCR was because of their higher sensitivities, although TaqMan qPCR also detected S. aureus-specific extracellular DNA.
Conclusion
This first evaluation of the GeneXpert MRSA/SA ETA assay with ETA samples found it to be highly sensitive, specific, user-friendly (hands-on time ~ 5 min.), and rapid (~ 66 min. assay time). Where this equipment is not available, we recommend implementing more sensitive culture-based methods for improved S. aureus detection in ETA samples.